Aerotech ARDS 150 Turntable Auto Tune/Calibration
- Install the Aerotech Soloist software supplied on the DVD with the turntable system.
- Place the device under test on the Turntable, ensure all cables are clear as the Turntable will rotate during calibration.
- Open the Soloist Digital Scope Software Module.
- Connect – navigate to the controller dropdown menu on the top toolbar, this contains the ‘connect’ option.
- Once connected Enable – do this by pressing the ‘enable’ symbol on the second toolbar, see image below.
When enabled the icon should change and display as green.
- Auto Tune – navigate to the ‘Action’ dropdown menu and select ‘auto tune’.
- In the auto tune window, enter 30 into the amplitude box.
- Press ‘Ok’, wait for the turntable to complete and results to be displayed.
- Accept the ‘Auto tune Results’
- Results – the ‘Pos Err’ values should be as low as possible and ideally below 0.2. You may need to run the auto tune more than once – if so, repeat steps 4-7 a couple of times.
-
When the turntable runs smoothly save the configuration, do this by pressing the ‘Commit Parameters’ button.